Easy Maintenance Low Service Cost
Optional Features
EtCO2, FiCO2 can be accurately measured. The performance of the partial CO2 rebreathing technique can be improved through means of arterial blood gas sampling and an algorithm that takes into account the effects of non-equilibration of EtCO2 during rebreathing.
Multiple Options for Different User
System Specifications
Pneumatic Ventilator | Operating Conditions | |||||
|
Ventilator: 455mm x 400mm x 300mm Trolley: 565mm X 535mmX 910mm
Ventilator: 12.5 kg Trolley: 20 kg |
|||||
Standards | ||||||
|
||||||
Alarm | Upper limit | Lower limit | ||||
Tidal volume Minute volume Oxygen concentration Airway pressure Respiratory frequency ETCO2 Pulse rate Circuit occlusion Apnea alarm O2 SUPPLY DOWN alarm Mains failure alarm Battery power low alarm Battery discharged alarm Alarm silence |
20~3000 mL 1~99 L 22~100%, OFF 1~99 cmH2O 1~120 bpm 0.1~13.3 kPa 31~250 bpm (PEEP+15) cmH2O, 15s 15~60s < 280 kPa Automatically switch < 10 min < 5 min ≤120 s |
OFF,20~2999 mL 0~98 L 20~100% 0~98 cmH2O OFF,1~99 bpm 0~13.2 kPa 30~249 bpm (PEEP+15) cmH2O, 15s 15~60s < 280 kPa Automatically switch < 10 min < 5 min ≤120 s |
||||
Setting | Parameter | Monitoring | Parameter | |||
Tidal volume Respiratory frequency Inspiratory time Breath-hold time Pressure triggering Flowrate triggering PEEP Pressure support Pressure control High-pressure level Low-pressure level Oxygen concentration 100% O2 T-high T-low freeze Expiratory hold Inspiratory hold Manual inspiration |
20~2500mL 1~120 bpm 0.1~12 s 0~50% --20cmH2O~PEEP 0 0.5~20 L/min 0~50 cmH2O 0~80 cmH2O 5~80 cmH2O 5~80 cmH2O 0~50 cmH2O 21~100% 90mins 0.1~30s 0.3~30s Yes Yes Yes Yes |
Inspiratory tidal volume Expiratory tidal volume Minute volume Spontaneous minute volume Spontaneous respiration frequency PEEP Airway peak pressure Mean airway pressure Inspiratory platform pressure Minimum airway pressure Inhaled oxygen concentration Airway resistance
Compliance RSBI PEEPi EtCO2 |
0~4000 mL 0~4000 mL 0~99 L 0~99 L/min 0~120 bpm 0~120 bpm 0~100 cmH2O 0~100 cmH2O 0~100 cmH2O 0~100 cmH2O -20~100 cmH2O 15% ~100 % 0~200 cmH2O/(L/S) 0~300 mL/cmH2O 0 ~ 9999 bpm/L 0~100cmH2O 0 ~ 100 mmHg |