Send Message

R55 Ventilator Machine For Hospital Tidal Volume setting 20-2500mL

R55 Ventilator Machine For Hospital Tidal Volume setting 20-2500mL
Features Gallery Product Description Request A Quote
Features
Specifications
Name: Medical Portable Touchscreen Siriusmed Ventilator
Keyword: Siriusmed Ventilator
Power Source: Electric
Material: Metal, Plastic
Warranty:: 1 Year
Instrument Classification: Class III
Widely Usage Range: Suitable For Infants And Adults
Tidal Volume Setting: 20~2500mL
Sufficient Modes Of Ventilation:: VCV, PCV, SIMV-VC, SIMV-PC, SPONT/CPAP+PSV, DuoPhasic
Friendly User Interface:: TFT Color Screen With Touch Screen
High Light:

R55 Ventilator Machine For Hospital

,

2500mL Ventilator Machine For Hospital

,

20ml hospital ventilator machine

Basic Infomation
Brand Name: R55
Payment & Shipping Terms
Payment Terms: T/T
Product Description

Easy Maintenance Low Service Cost

  • Friendly interface, easy to cooperate. Low costs of spare parts, no more tie up of funds.
  • Easy maintenance by cartridge type, easy replacement of service parts.
  • Simple to maintenance makes an efficient daily routine. Software upgrade by USB.
  • Rapid respond after sales service, save your time, reduce the risks.

Ventilator R55-Easy to position ergonomic trolley with two brakes,HFNC.

Multiple Options for Different User

  • High-end model R55 have more advance modes of ventilation, with the high-end functions such as PRVC, NIV, VS. It will be easily updated via software.
  • With the optional modules of famous brands, such as EtCO2 nebulizer and with more monitoring information for doctor’s reference.
  •  

System Specifications

Pneumatic Ventilator Operating Conditions
  • 12",TFT display, with touch screen
  • VCV, PCV, SIMV-VC, SIMV-PC, CPAP/PSV,

(Optional PRVC,NIV,VS) :见上描述

,BACKUP

  • P-T, F-T, V-T

CO2-time, pulse rate-time,

loop: P-V, F-V ,P-F

  • NI-MH battery, >90mins
  • RS232, VGA, USB, Remote alarm
  • Gas supply: Air, O2, 280kPa -600kPa (41~87 PSI)
  • Power supply: 100-240V, 50/60Hz
  • Temperature: 10-40℃ (Operation),
  • ℃ (Storage)
  • Relative humidity: 5%-95%,

non-condensing (Operation)

  •  
  • Atmospheric pressure

(Operation & Storage): 500hPa-1060hPa

  • Dimensions (H×W×D):

Ventilator: 455mm x 400mm x 300mm

Trolley: 565mm X 535mmX 910mm

  • Weight:

Ventilator: 12.5 kg

Trolley: 20 kg

Standards
  • IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-12
Alarm Upper limit Lower limit

Tidal volume

Minute volume

Oxygen concentration

Airway pressure

Respiratory frequency

ETCO2

Pulse rate

Circuit occlusion

Apnea alarm

O2 SUPPLY DOWN alarm

Mains failure alarm

Battery power low alarm

Battery discharged alarm

Alarm silence

20~3000 mL

1~99 L

22~100%, OFF

1~99 cmH2O

1~120 bpm

0.1~13.3 kPa

31~250 bpm

(PEEP+15) cmH2O, 15s

10~60s

< 280 kPa

Automatically switch

< 10 min

< 5 min

≤120 s

OFF,20~2999 mL

0~98 L

20~100%

0~98 cmH2O

OFF,1~99 bpm

0~13.2 kPa

30~249 bpm

(PEEP+15) cmH2O, 15s

10~60s

< 280 kPa

Automatically switch

< 10 min

< 5 min

≤120 s

Setting Parameter Monitoring Parameter

Tidal volume

Respiratory frequency

Inspiratory time

Breath-hold time

Pressure triggering

Flowrate triggering

PEEP

Pressure support

Pressure control

High pressure level

Low pressure level

Oxygen concentration

100% O2

T-high

T-low

freeze

Expiratory hold

Inspiratory hold

Manual inspiration

20~2500mL

1~120 bpm

0.1~12 s

0~50%

-20cmH2O[1] ~PEEP 0

0.5~20 L/min

0~50 cmH2O

0~80 cmH2O

5~80 cmH2O

5~80 cmH2O

0~50 cmH2O

21~100%

2mins

0.1~30s

0.3~30s

Yes

Yes

Yes

Yes

Inspiratory tidal volume

Expiratory tidal volume

Minute volume

Spontaneous minute volume

Spontaneous respiration frequency

PEEP

Airway peak pressure

Mean airway pressure

Inspiratory platform pressure

Minimum airway pressure

Inhaled oxygen concentration

Airway resistance

 

Compliance

RSBI

PEEPi

EtCO2

0~4000 mL

0~4000 mL

0~99 L

0~99 L/min

0~120 bpm

0~120 bpm

0~100 cmH2O

0~100 cmH2O

0~100 cmH2O

0~100 cmH2O

-20~100 cmH2O

15% ~100 %

0~600 cmH2O/(L/S)

0~300 mL/cmH2O

0 ~ 9999 bpm/L

0~100cmH2O

0 ~ 100 mmHg


0-20cmH2O

  • Pneumatically driven and electronically controlled ICU ventilator
  • Widely usage range: Suitable for infants and adults, newborns
  • Sufficient modes of Ventilation: VCV, PCV, SIMV-VC+PSV, SIMV-PC+PSV, SPONT/CPAP+PSV, BACKUP, Manual (Optional for PRVC,VS,NIV-PC,NIV-CPAP, DuoPhasic+PSV ,APRV+PSV,VS)
  • Knob and Hard key input,Easy and fast choose functions and one step start
  • Easy to position ergonomic trolley with two brakes
  • Quick self-test no need frequent interaction from operator
  • Patient settings automatic stored and can be restored quickly
  • Direct setting of ventilation parameters in the main interface
  • 3 waveforms and loops on one interface, quick changeover
  • System log can record 100 alarms

 

 